    Contents

    I. SUMMARY.............................................................................

    II. MAINSPECIFICAITON....................................................................................- 2 -

    III. STRUCTURE....................................................................................................- 4 -

    IV. INSTALLATION................................................................................................- 5 -

    V. OPERATION......................................................................................................- 7 -

    VI. PROBLEMS ANDTROUBLESHOOTING................................................. -14 -

    VII.MAINTENANCE............................................................................................- 14 -

    VIII. CAUTION......................................................................................................- 15 -

    IX. APPENDIX......................................................................................................- 16 -

    X. REFERENCE FOR TEMPERATURE VALUE............................................ - 16 -

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    3/20

    Warning: The instrument can not provide the designed protectionforoperators who do not follow the right procedures andrequirements given by

    the manufacturer.

    Warning: All solutions must be handled with care according tothe

    labs safety regulation. Please make a reference to the relatedmaterial safety

    data sheet. Wear the lab-gown, goggle and rubber gloves all thetime. Be

    care of hot reagents.

    Warning: Be aware of the risk of electric shock. Only thetrained

    professionals are permitted to open the face panel or backcover.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    4/20

    )$46 User Manual

    - 1 -

    I. Summary

    1. Application

    )$46 Fat Analyzer, designed in accordance with GB/T 14772-2008,can

    be used not only in measuring the fat content in food, diaryproduct and

    feedstuff, but also in testing the soluble organic compound indetergent,

    medicine, petrochemicals, fiber products, soil and sewageetc.

    For example:

    (1). To test the fat content in the food, feedstuff, grain, andseed.

    (2). To extract the fat from the sewage and mud.

    (3). To extract the semi-volatile organic compound such aspesticide and

    weedicide.

    (4).To extract plasticiser from plastics; to extract colophonyfrom the paper; to

    extract grease from the leather.

    (5).To perform the pre-digestion process to the solid samplesfor the Gas

    chromatography (GC) or Liquid chromatography (LC).

    2. Principle

    Based on the principle of Soxhlet extraction, )$46 uses theweighing

    method to obtain the fat content. After being extracted from thesample by

    solvents such as absolute ether or petroleum ether, the fat isdried and

    weighed. The difference between the vessels weights before andafter the

    extraction is the fat content.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    5/20

    - 2 -

    II. Main Specificaiton

    1. Technical specification

    (1). Wide temperature controlling range:

    Room temperature: +5~280; Accuracy: 1

    (2). Heating power: 1000 W

    (3). Operation voltage: 220V(4). Frequency: 50/60 Hz

    (5). As many as six samples can be tested at the same time.

    (6). Volume of solvent vessel: 80 ml;

    Sample amount: 0.5~15 g (regular amount: 2~5 g)

    (7). Solvent recovery rate: 80%

    (8). Testing scope: 0~100%

    (9). Dimension (mm): 650320700

    (10). Weight: 35 Kg

    2. Features:

    (1).5.1 inch LCD and controlling system with microprocessor

    (2).Overheating warning and alarm function are available.

    (3).The audible, visual warning signals as well as the remindingmessages on

    the screen are used in the alarming system.

    (4).Both timing and time recording functions are available.

    (5).The all-in-one metal heating block is utilized to achievethe wide

    temperature controlling range with the high controllingaccuracy.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    6/20

    - 3 -

    (6).The elevation connection using the transmission technique ofthe linear

    bearing brings the flexible elevating operation.

    (7).All information can be clearly displayed on the screen. Thepreset

    temperature, actual temperature and heating time can bedisplayed

    simultaneously on the screen.

    (8).The safe operation is ensured because the electric circuitis completely

    isolated with the extraction cell.

    (9).With the unique air space insulation technique, thetemperature ofinstruments case can be kept normal.

    (10). Compared with the traditional soxhlet extraction, itsspeed and

    efficiency are better.

    3. Working Condition

    (1). Input voltage: 220V 50Hzair break switch and leakagecircuit breaker

    are required.

    (2). The apparatus should be placed where water, drain basin,electric pocket

    are available.

    (3). Power supply should meet the requirements of the apparatusto avoid

    electric over stress; independent switch, safety device andreliable grounding

    are required.(4). The apparatus should be placed away from bigelectric equipment to

    avoid strong magnetic field.

    (5). Ensure the good ventilation of the laboratory.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    7/20

    - 4 -

    III. Structure

    Components

    1.Gliding balls 2. Condenser pipe 3. co*ck 4. Extraction cups

    5. Lever 6. Socket 7. Power switch

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    8/20

    - 5 -

    IV. Installation

    1. Check before Installation

    After the analyzer is unpacked, check the analyzer and all theparts specified

    in the manual and check if there are any damages. If so, pleasecontact the

    manufacturer in time (please keep the damaged parts).

    2. Conditions for Installation

    (1). This analyzer should be placed away from direct sunshine,moisture,

    heat and coldness,Generally room temperature should be keptbetween

    10 ~ 28.

    (2). This analyzer should be placed near water source and sinks,and electric

    socket should be available at the site. The distance between theinstrument

    and water tap (or the power socket) should not be over 1 meterto ensure

    convenience of operation. The water should meet the followingrequirements:

    water flow 1-2 liter per minute, adequate water pressure andtemperature no

    more than 20.

    (3). The sink should be at least 10cm lower than the wateroutlet of the

    instrument to ensure free flow of the water.

    (4). The power supply should meet the requirement. Ground line,separate

    power switch and right fuse are necessary to make sure thesafety of the

    operators.

    (5). This analyzer should be installed at the place where thereis no big

    electrical equipment nearby and the working place should be freeof vibration,

    corrosive liquid and interference from strong electromagneticfield.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    9/20

    - 6 -

    3. Procedures of Installation

    Put the analyzer at test table stably. The back of it should be20 centimeters

    away from the wall, and the socket should not be more than 1meter away

    from the analyzer. Air switch, leakage switch, and reliableground line

    should be equipped. See figure 1.

    Figure 1

    (1). Condensate inlet (connect to the water tap)

    (2). Condensate outlet (connect to the sink)

    )$46 User Manual

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    10/20

    - 7 -

    V. Operation

    1. Reagent, equipment and tools

    (1). Ether (analytical pure), petroleum ether (analytical pure,boiling range

    30~60)

    (2). Analytical balance (0.0001g)

    (3). Filter paper with diameter of 90mm

    (4).Graduated cylinder (100ml)

    (5). Drying oven

    (6). Grinder

    (7). Mortar and pestle

    (8). Desiccator

    2. Pre-processing of Sample

    2.1 Pre-processing of Food Sample

    2.1.1 Solid sample: Get at least 200g of typical sample. Mashand mix the

    sample well, and put it into the closed glass vessel; for thosethat are not

    easy to mash, cut them into small granules and then put theminto the closed

    glass vessel through a circular sieve ( holes 1.0 mm indiameter).2.1.2 Powder sample: get at least 200g of typical sample(the powder should

    be mashed well if particles of the powder are big). Mix well,and then put into

    glass vessel through a circular sieve ( holes 1.0 mm indiameter).

    2.1.3 Paste sample: get at least 200g of typical sample, mixwell, and put into

    glass vessel.

    2.1.4 Solid and liquid sample: get at least 200g of typicalsample according to

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    11/20

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    12/20

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    13/20

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    14/20

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    15/20

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    16/20

    - 13 -

    4. Experiment Procedures

    (1). Make one extraction thimble with filter paper (use thepaper rod attached).

    Measure 2-5g sample with the analytical balance (to the nearest0.0001g)

    and put the sample in the thimble. Put the thimble in theholder, which will be

    attached to the instrument by magnet.

    (2). Dry a constant weight extraction cup and measure itsweight. Put 60ml

    Petroleum ether or absolute ether in the cup and put it on theheating plate.

    (3). Push down the lever to seal the extraction cup with thejoint washer.

    Move the gliding ball to put the thimble in the cup and makesure that the

    sample is fully in the solvent.

    (4). Turn on the power and open the condensate. Set up thetemperature and

    time.

    (5). After a while, move the gliding ball to lift up the thimbleand start the

    extraction. When the extraction is completed, close the co*ck onthe

    condensate pipe and the solvent will be evaporated, condensed atthe

    condenser, and collected in the empty extraction chamber underthecondensate pipe. Then open the co*ck and start the rinsingprocess, during

    which the condensed solvent washes final traces of fat from thesample. In

    the end, close the co*ck and the solvent will be condensed andcollected in

    the extraction. Stop heating.

    (6). Cool the heating plate and extraction cup. Pull up thelever and open the

    co*ck. Get the solvent in the extraction chamber.

    (7). Remove the extraction cup from the heating plate and put itin the drying

    oven. Then put the cup in the desiccator to cool down. Measurethe weight of

    the cup and work out the percentage of the fat content.

    (8). Turn off the power and condensate.

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    17/20

    - 14 -

    VI. Problems and Troubleshooting

    No. Problem Possible Reason Trouble Shooting

    1 No powerThe fuse burnedBad connection of powercord

    Replace the fusePlug the power cordagain

    2SolventGasleakage

    1.Insufficient condensate

    2.The rubber ring at the end

    of the condensate pipe is old

    and can not seal the pipe

    tightly

    3.The condensate pipe is

    broken

    Put in more condensateReplace the rubberringReplace thecondensatepipe

    3Systemfailure orcrash

    Check if there is any strongmagnetic field or electronicfieldaround

    Turn off theinstrument,move awaythe source ofdisturbance andrebootthe system

    VII. Maintenance

    1. The analyzer needs heating up, so good ventilation and heatdissipationare necessary.2. It is recommended that in the certainperiod, the same extraction systemshould use the same solvent toprevent cross-contamination betweendifferent solvents.3. Clean theinstrument with a wet cloth to avoid static electricity; cleantheheating plate with an abrasive cloth to improve heattransmission.

    4. Keep the receiving flasks and extraction containers clean.Clean them

    )$46 User Manual

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    18/20

    - 15 -

    regularly.5. Make sure that there is no seaweed growing in thecooler. If there is, cleanthe cooler with hydrochloric acid orhypochlorous acid (1 mol/L).

    VIII. Caution

    1. Read the manual carefully and follow the instructions. Theinstrument must

    not be operated by any green hand.

    2. The instrument must be cleaned with a wet cloth, especiallythe glass

    wares. A dry one may generate static electricity and causedanger.

    3. The maintenance must follow the instructions strictly.Necessary

    replacement of glass wares, parts and wires can be done underthe

    instruction of the professtional engineers. But other repairwork must be done

    by the professtional engineers.

    4. We does not guarantee safety after any unauthorizedalteration is done to

    the instrument. Therefore, we does not take responsibility forany loss

    thereafter.

    5. Cut off the power before opening the instrument and replacingthe fuse

    pipe.

    6. Handle with care. There are some glass wares in theinstrument.7. If the instrument will not be used for a long time,remove the solvent in the

    condensate pipes.

    8. It is recommended that in the certain period, the sameextraction system

    should use the same solvent to prevent cross-contaminationbetween

    different solvents.

    9. Mismanagement of organic solvent may cause explosion, so keepthe

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    19/20

    - 16 -

    organic solvent away from fire and static electricity.

    10. Dont touch the instrument when its heating up.

    11. Dont leave the instrument in experiment unattended.

    IX. Appendix

    This product is covered by the one-year limited warranty sincethe day of

    purchase. The limited warranty does not cover the followingsituations:

    1. The warranty expired.

    2. The instrument is damaged by any inappropriate use.

    3. The instrument is damaged by any unauthorizeddisassembly.

    4. The instrument is damage by any improper storage.

    X. Reference for Temperature Value

    The heating temperature should be adjusted according to theroom

    temperature and the heat transmission efficiency from theheating plate to the

    extraction cups. The dipping rate will be about 3-5dips/ secondin proper

    temperature. The following chart can be referred to for theextraction

    temperature of some common organic solvents.

    )$46 User Manual

  • 7/28/2019 Manual Operacion Del Equipo Analizador Fa46

    20/20

    - 17 -

    SolventExtraction Temperature

    30-60 Petroleum

    ether80

    Ether 60

    Acetone 76

    methylene dichloride 60

    Hexane 89

    Methanol 85Ethanol 98

    chloroform 82

    benzene 100

    Attention: for the organic solvent not listed above, theextraction

    temperature can be 20

    higher than the boiling point.

Manual Operacion Del Equipo Analizador Fa46 - [PDF Document] (2024)
Top Articles
Latest Posts
Article information

Author: Lidia Grady

Last Updated:

Views: 5886

Rating: 4.4 / 5 (65 voted)

Reviews: 80% of readers found this page helpful

Author information

Name: Lidia Grady

Birthday: 1992-01-22

Address: Suite 493 356 Dale Fall, New Wanda, RI 52485

Phone: +29914464387516

Job: Customer Engineer

Hobby: Cryptography, Writing, Dowsing, Stand-up comedy, Calligraphy, Web surfing, Ghost hunting

Introduction: My name is Lidia Grady, I am a thankful, fine, glamorous, lucky, lively, pleasant, shiny person who loves writing and wants to share my knowledge and understanding with you.